Types of equipment and Anaesthesia Machine

This guide integrates essential standards for gas supply, circuits, airway management, and protective ventilation strategies.


1. Gas Supply, Safety, and Vaporizers

Pin Index Safety System (PISS) prevents the connection of the wrong gas cylinders.

Gas Colour (Int’l) Pin Index
Oxygen White / Green 2, 5
Nitrous Oxide Blue 3, 5
Air Grey / White 1, 5

Vaporizer Colour Coding:

  • Isoflurane (Purple),
  • Sevoflurane (Yellow),
  • Desflurane (Blue).

2. Breathing Circuits & Airway Management

  • Mapleson Circuits: Classified by FGF, bag, and APL valve arrangement. Mapleson D (Bain circuit) is the most common clinical coaxial system.
  • ETT Types: PVC (standard), Reinforced (wire-coiled, avoids kinking), Laser-resistant (for airway surgery), and Double-lumen (independent lung ventilation).
  • LMA: Supraglottic device, fundamental for both routine use and “cannot intubate, cannot oxygenate” (CICO) rescue algorithms.
  • ASA Difficult Airway Protocol: Emphasizes preparation: Videolaryngoscopy, Bougie, Fiberoptic scope, and predefined rescue plans.

3. Ventilation Modes

Mode Technical Description
VCV Volume Control: Delivers a fixed tidal volume at a set rate.
PCV Pressure Control: Delivers flow to a set inspiratory pressure.
SIMV Synchronized Intermittent Mandatory Ventilation: Combines machine breaths with patient-triggered breaths.
PSV Pressure Support: Entirely patient-triggered; assists each spontaneous breath.


Critical Care Note: Lung-Protective Strategy

To prevent VILI (Ventilator-Induced Lung Injury), utilize:

  • Low Tidal Volume: 6 ml/kg (Predicted Body Weight).
  • PEEP Optimization: To maintain alveolar recruitment without over-distension.

For more high-yield medical resources and free medical questions, visit mymedschool.org.